Pt100, Pt500 and Pt1000 calculator: resistance to temperature

Convert resistance to temperature and temperature to resistance for Pt100, Pt500 and Pt1000 RTDs (IEC 60751), Ni1000, NTC and thermocouples (mV ⇄ °C).

Fill in 1 of the 2 fields and the missing one is calculated. You can type in any of them.

Resistance of the sensor; for thermocouples, the measured voltage.

Use this calculator for reference only, as we cannot guarantee that the results are free of errors.

Disclaimer

https://dyaii.es/en/pt100-resistance-to-temperature/

Advertisement

Pt100, Pt500 and Pt1000 sensors are platinum resistors whose value changes with temperature. This calculator converts temperature to resistance and resistance to temperature according to IEC 60751, from −200 to 850 °C. It also works for nickel sensors (Ni100 and Ni1000), copper sensors (Cu100), NTC thermistors and type K, J, T, E, N, R, S and B thermocouples, which give a voltage in millivolts.

How to use the calculator

  1. Choose the sensor: Pt100, Pt500, Pt1000, Ni100, Ni1000, Cu100, NTC or a thermocouple.
  2. Type the temperature and you get the resistance (or the thermocouple voltage), or type the measured resistance (or voltage) and you get the temperature.
  3. For an NTC, enter its nominal resistance R₀, its reference temperature T₀ (almost always 25 °C) and its β constant.
  4. For a thermocouple, enter the cold junction temperature (that of the terminals where it is measured). At 0 °C you get the voltage in the tables.
To measure with a multimeter, disconnect the sensor from the instrument. With 2 wires, the cable resistance is added to the reading.

Callendar-Van Dusen equation

IEC 60751 defines the sensor resistance (R₀ = 100, 500 or 1000 Ω at 0 °C) as:

R(T) = R₀ · (1 + A·T + B·T²)   if T ≥ 0 °C
R(T) = R₀ · (1 + A·T + B·T² + C·(T − 100)·T³)   if T < 0 °C

With A = 3.9083 × 10⁻³, B = −5.775 × 10⁻⁷ and C = −4.183 × 10⁻¹². To go from resistance to temperature, the equation is solved the other way around.

Advertisement

Pt100, Pt500 and Pt1000 resistance table

Temperature (°C)Pt100 (Ω)Pt500 (Ω)Pt1000 (Ω)
−20018.5292.60185.20
−10060.26301.28602.56
−5080.31401.53803.06
−2092.16460.80921.60
0100.00500.001000.00
20107.79538.971077.93
25109.73548.671097.35
50119.40596.991193.97
100138.51692.531385.05
150157.33786.631573.25
200175.86879.281758.56
300212.051060.262120.51
400247.091235.462470.92
500280.981404.892809.78
600313.711568.543137.08
850390.481952.413904.81

Tolerance classes

ClassTolerance
AA±(0.1 + 0.0017 · |t|) °C
A±(0.15 + 0.002 · |t|) °C
B±(0.3 + 0.005 · |t|) °C
C±(0.6 + 0.01 · |t|) °C

Ni100 and Ni1000 nickel sensors

Nickel sensors change more per degree than platinum ones (α = 0.00618 versus 0.00385), but their range is narrower and their curve deviates more from a straight line. They are used mainly in HVAC. DIN 43760 gives their resistance from −60 to 250 °C:

R(t) = R₀ · (1 + A·t + B·t² + D·t⁴ + F·t⁶)

With R₀ = 100 Ω (Ni100) or 1000 Ω (Ni1000), A = 5.485 × 10⁻³, B = 6.65 × 10⁻⁶, D = 2.805 × 10⁻¹¹ and F = −2 × 10⁻¹⁷.

Cu100 copper sensor

Copper is cheap and almost linear. With α = 0.00428 (GOST 6651), from 0 to 200 °C the resistance is a straight line, and below 0 °C (down to −180 °C) it has a small correction:

R(t) = R₀ · (1 + A·t)   if t ≥ 0 °C
R(t) = R₀ · (1 + A·t + B·t·(t + 6.7) + C·t³)   if t < 0 °C

With A = 4.28 × 10⁻³, B = −6.2032 × 10⁻⁷ and C = 8.5154 × 10⁻¹⁰.

Temperature (°C)Ni100 (Ω)Ni1000 (Ω)Cu100 (Ω)
−6069.52695.2074.10
−5074.26742.5578.46
0100.001000.00100.00
20111.241112.36108.56
25114.131141.29110.70
50129.101291.05121.40
100161.781617.79142.80
150198.631986.35164.20
200240.662406.60185.60
250289.162891.56—

NTC thermistors: the β equation

An NTC is a semiconductor whose resistance drops sharply as it heats up. The manufacturer gives the nominal resistance R₀ at a reference temperature T₀ (almost always 10 kΩ at 25 °C) and the β constant (B25/85 or B25/50), and with them:

R(T) = R₀ · eβ · (1/T − 1/T₀)
T = 1 / (1/T₀ + ln(R/R₀) / β)

With the temperatures in kelvin (T = t + 273.15). The β equation is exact at T₀ and at the other temperature used to calculate β; between −40 and 125 °C it deviates a few tenths of a degree from the real curve. If you need more accuracy, use the manufacturer’s table.

Temperature (°C)NTC 10 kΩ, β = 3950 K (Ω)
−40401,860
−20105,385
033,621
1020,175
2012,535
2510,000
308037
405301
503588
602486
801270
100698
125359

Example: a 10 kΩ NTC with β = 3950 K reads 5 kΩ. T = 1 / (1/298.15 + ln(0.5) / 3950) = 314.61 K, that is, about 41.5 °C.

Thermocouples: voltage and temperature

A thermocouple is two wires of different metals welded together at the tip (the hot junction). Between their free ends (the cold junction) a voltage of a few millivolts appears that depends on the temperature difference between the two junctions. The relationship between voltage and temperature for each type is given by the NIST ITS-90 reference functions, the same as in IEC 60584-1, with the cold junction at 0 °C:

E(t) = Σ cᵢ · tⁱ

Each type has its own coefficients cᵢ for each temperature range. Type K, above 0 °C, also adds a₀ · ea₁ · (t − a₂)², with a₀ = 0.1185976 mV, a₁ = −1.183432 × 10⁻⁴ °C⁻² and a₂ = 126.9686 °C.

To go from voltage to temperature, the calculator uses the NIST inverse polynomials (t = Σ dᵢ · Eⁱ, with an error below 0.06 °C). Where NIST gives no inverse polynomial (below −200 °C for K, T, E and N, and below 250 °C for B), it solves the reference function the other way around.

Temperature (°C)K (mV)J (mV)T (mV)E (mV)N (mV)R (mV)S (mV)B (mV)
−200−5.891−7.890−5.603−8.825−3.990———
−100−3.554−4.633−3.379−5.237−2.407———
000000000
251.0001.2770.9921.4950.6590.1410.143−0.002
1004.0965.2694.2796.3192.7740.6470.6460.033
2008.13810.7799.28813.4215.9131.4691.4410.178
40016.39721.84820.87228.94612.9743.4083.2590.787
60024.90533.102—45.09320.6135.5835.2391.792
80033.27545.494—61.01728.4557.9507.3453.154
100041.27657.953—76.37336.25610.5069.5874.834
120048.83869.553——43.84613.22811.9516.786
1400—————16.04014.3738.956
1600—————18.84916.77711.263
1800———————13.591

Thermocouple types: metals, colors and ranges (IEC 60584)

The colors are those of IEC 60584-3 for extension and compensating cables: the jacket and the positive wire have the color of the type, and the negative is always white. The ranges are those of the reference functions; in practice, each thermocouple is used over a narrower range depending on the wire diameter and the sheath.

TypePositive (+)Negative (−)IEC colorRange (°C)Sensitivity from 0 to 100 °C
KNickel-chromiumNickel-aluminumGreen−270 to 137241 µV/°C
JIronCopper-nickelBlack−210 to 120053 µV/°C
TCopperCopper-nickelBrown−270 to 40043 µV/°C
ENickel-chromiumCopper-nickelViolet−270 to 100063 µV/°C
NNickel-chromium-siliconNickel-siliconPink−270 to 130028 µV/°C
RPlatinum-13 % rhodiumPlatinumOrange−50 to 1768.16.5 µV/°C
SPlatinum-10 % rhodiumPlatinumOrange−50 to 1768.16.5 µV/°C
BPlatinum-30 % rhodiumPlatinum-6 % rhodiumGray0 to 18200.3 µV/°C
Type B gives almost no voltage below 50 °C (between 0 and 40 °C it is around 0 mV): that is why it needs no cold junction compensation, but it is also useless for measuring low temperatures.

Cold junction compensation

A thermocouple does not measure the temperature at the tip, but the difference with that of its terminals. If the terminals are at Tf, the measured voltage is:

E measured = E(T) − E(Tf)   →   E(T) = E measured + E(Tf)

Transmitters and PLCs measure the temperature of the terminals and add E(Tf) automatically. If you measure the millivolts with a multimeter, add the voltage for the ambient temperature yourself before looking it up in the table, or enter it in “Cold junction temperature”. The cable between the thermocouple and the instrument must be an extension or compensating cable of the same type: with copper cable, the cold junction moves to the junction box.

Worked example with a type K thermocouple

A multimeter reads 8.5 mV at the terminals of a type K thermocouple, and the terminals are at 22 °C:

  1. Cold junction voltage: E(22 °C) = 0.879 mV.
  2. Voltage referred to 0 °C: 8.5 + 0.879 = 9.379 mV.
  3. With the type K inverse polynomial: T ≈ 230.9 °C.

Without compensation (looking up 8.5 mV in the table) you would get about 209 °C: an error of more than 20 °C.

Advertisement

Frequently asked questions

What is the difference between Pt100 and Pt1000?

A Pt1000 has ten times more resistance, so the lead wire resistance has ten times less effect. That is why it is preferred with 2-wire connections or long cables.

2, 3 or 4 wires?

With 2 wires the cable resistance is added to the reading. With 3 wires it is compensated if the leads are identical, and with 4 wires it is eliminated completely: it is the most accurate connection.

What is thermocouple cold junction compensation?

A thermocouple measures the temperature difference between the tip and the terminals where it joins the copper. The instrument measures the temperature of those terminals and adds its voltage: E(T) = E measured + E(Tf). Enter that temperature in “Cold junction temperature”.

Which beta should I use for an NTC?

The one on its datasheet (B25/85 or B25/50). The most common 10 kΩ NTCs have β between 3435 and 3950 K. The β equation is exact at the two temperatures it was calculated from and drifts somewhat outside them.

Which thermocouple should I choose?

Type K is the most widely used (from −200 to 1260 °C in practice); type J, for furnaces without oxygen up to about 750 °C; type T, for low temperatures; type N, more stable than K at high temperature; and R, S and B, made of platinum, for over 1000 °C.